1-4 5000 Multi Point Moisture Analyzer
Gas Flow (Figure 2)
Gas flows in three separate streams through the system: sample gas, reference gas, and moisture
generator gas. Flow rate through the cell (for whichever gas is present) is controlled by flow control
needle valves and indicated by a flowmeter at the cell exhaust. A pressure gauge is also located at
the exhaust port to ensure analyzer operation at a back pressure of 103 kPa (15 psi) gauge; this per-
mits exhausting to a pressurized vent.
Four solenoid valves (L
1-4
) are switched by the controller to control gas flow to the cell as described
below. Note that energized valves are in the normally closed contacts (NC) position and de-ener-
gized valves are in the normally open contacts (NO) position:
Normal Operation:
Sample gas flow (ƒ
sample
measurement)
L
1
is energized and L
4
is de-energized so sample gas flows to the cell. L
2
and L
3
are de-
energized (open contacts) so reference and moisture generator gases are exhausted.
Reference gas flow (ƒ
ref
measurement)
L
3
and L
4
remain de-energized. L
1
and L
2
toggle such that reference gas flows through the
cell and sample gas is exhausted.
Auto or Manual Verification:
Moisture Generator gas flow (ƒ
cal
measurement)
L
3
and L
4
are energized so sample bypasses L
1
to exhaust port and calibration gas flows to L
1
. L
1
is energized so moisture generator gas flows to the cell. L
2
is de-energized so reference gas is
exhausted.
Reference gas flow (ƒ
ref
measurement)
L
3
and L
4
remain energized. L
1
and L
2
toggle such that reference gas flows through the cell and
moisture generator gas is exhausted.
System Components
A typical 5000 Moisture Analyzer system is comprised of an explosion-proof 5000 field unit, sample
drying and filtering equipment, and a remote 5000 Microprocessor Controller.
Field Unit (Figure 3)
Components contained in the  eld unit:
Sample cell contained in an oven and maintained at 60°C
Sample and reference solenoid valves (L
1
, L
2
)
Moisture generator
Moisture generator solenoid valves (L
3
, L
4
)
Power and electronics control card
Power transformer (not shown)
Internal heater and 40°C thermostat (not shown)
H2S scrubber (optional, not shown)