2-2 | Heated Acid Gas Sample Probe (Large Diameter)
IECEx: IECEx Certificate No.:
IECEx DEK 15.0035X;
Ex db IIB+H
2
T3 Gb or
Ex db IIB+H
2
T3
GOST: 1ExdIIBT3
Complies with all relevant European Directives,
GOST Pattern Approval
RoHS: Restriction of Hazardous Substances Directive:
2011/65/EU (RoHS 2)
Ingress Protection
NEMA Type 4, IP 55
Ambient Limits
-29 °C to +70 °C (-20.2 °F to +158 °F),
typical operating conditions with AMETEK Insulating Jacket.
Cable/Cable Glands
This typically applies to European/Hazardous Locations only.
Heat resistant cable and Ex d cable glands must be used, suitable for 85 °C
(185 °F).
Probe Valve Assembly
Typical Flow Rate
2 L/min (0.7 SCFM)
Aspirator Drive Gas (Sample Transport)
Instrument air, nitrogen, or steam, depending on the process.
Flow Rate: <1.0 SCFM (30 L/min) at 30 PSIG (2.1 BARG)
above the process pressure.
Air Quality (if used): As per ANSI/ISA S7.0.01 (1996) Quality Standard
for Instrument Air.
NOTE