Page 20
User GuideLSPHD Linescanner Head
4.1 UsingtheLaserAlignmentUnit
TheLaserAlignmentUnitisaCLASS2laserproduct.
BeawareofallCLASS2LaserSafetyWarningswhenevertheLaserAlignmentUnitisin
use.
1) PresstheLaserAlignmentbutton.Thebuttonisnexttotheinterfaceconnector
onthetopofthescannerbody.
Thescanfrequencyoftheinstrumentreducestoapproximately1Hzandthe
LaserAlignmentbuttonashestoindicatethatthelaserisactive.Aredstripeof
laserlightisprojectedthroughthewindowofthescanner,indicatingtheposition
and extents of the scan line.
2) Adjustthescannerpositionasrequiredsothattheproductalmostllsthewidth
ofthescandenedbythelaser.
Note
WhentheLaserAlignmentUnitisactive,thescannerdoesnotmeasurethe
target temperature.
3) Toswitchoffthelaser,presstheashingLaserAlignmentbutton.
WhentheLaserAlignmentunitisswitchedoff,thescannerstabilisesfor
approximately60seconds.Thescannerwillthenbeginnormaloperationand
temperature information is available.
Alternatively,theLaserAlignmentfacilitycanbeoperatedremotelyfromtheLandscan