Newage Testing Instruments, Inc.
7
To advance, click NEXT to “Measurement” tab. The
operator may select to record averages of multiple
tests rather than the individual tests.
Enter a value
into the Group Size.
The buttons on the remaining
selections are enabled.
The selections are as
follows:
Keep
All
The default setting.
Uses all measurement.
Eliminate Highest and Lowest Values
When 3 or more tests are selected to be averaged
the system will remove the highest and lowest
values from the group average.
For example, if the
average is set for 5, then 3 results will go into the
average.
If
an
average
of
2
is
selected
the
average
function
has
no
effect.
Eliminate furthest from the average.
Once the testing is completed the system calculates the average and removes the one test that is furthest from that
average.
If an average of 2 is selected the average function has no effect
Eliminate if standard deviation exceeds [selection]
The operator is prompted to enter a maximum standard deviation value.
When the testing is done the value farthest from
the standard deviation allowable is removed.
If all the values fall within the standard deviation, then none are removed.
Set Tolerance; After
Average
,
you
can set a
Tolerances.
The
program
uses
these
settings
to show green for good reading
and Red for bad readings.