931003E / 0620 4. FIRMWARE COMMANDS AND RESPONSES
Responses:
$IT<CR> Zero stabilization is currently enabled.
$IF<CR> Zero stabilization is currently disabled.
SHOW_ZERO_WIDTH
Reports the current width for the stabilizer zero peak. See also SET_ZERO_WIDTH,
SET_ZERO_CHANNEL, and SHOW_ZERO_CHANNEL.
Responses:
$C00001088<CR> Zero width has not been set.
$C00003089<CR> Zero width is 3 channels (lowest possible width).
... ...
$C00256100<CR> Zero width is 256 channels
START [seg-mask]
Starts data acquisition. The optional segment mask is provided for compatibility with other
MCBs and may be any value from 0 to 65535 but is ignored by the digiBASE.
STOP [seg-mask]
Stops the data acquisition. The optional segment mask is provided for compatibility with
other MCBs and may be any value from 0 to 65535 but is ignored by the digiBASE.