11 
       
recognize that the instrument has been connected.  If the correct driver is not found, follow the 
Windows on-screen prompts and install the required files from the CD supplied. 
LAN (XPF60-20DP only) 
The LAN interface is designed to meet LXI ( Lan eXtensions for Instrumentation) version 1.2; the 
instrument is Class C compliant. Remote control using the LAN interface is possible using a 
TCP/IP Socket protocol.  The instrument also contains a basic Web server which provides 
information on the unit and allows it to be configured.  Since it is possible to misconfigure the LAN 
interface, making it impossible to communicate with the instrument over LAN, a LAN 
Configuration Initialize (LCI) mechanism is provided via a recessed switch on the rear panel 
(marked LAN RESET) to reset the unit to the factory default.  
Further details are given in the Remote Operation chapter.  For more information on LXI 
standards refer to   www.lxistandard.org/home 
GPIB (XPF60-20DP only) 
The GPIB signal grounds are connected to the instrument ground. The implemented subsets are: 
SH1 AH1 T6 TE0 L4 LE0 SR1 RL2 PP1 DC1 DT0 C0 E2 
The GPIB address is set from the front panel.