Technical data
SPECTRO ARCOS FHX3X — 08.02.2022 — 23
4 Technical data
Description Value
Instrument height 1068 mm; 42.1"
Instrument width 1582 mm; 62.3"
Instrument depth 756 mm; 29.8"
Instrument weight approx. 249 kg; approx. 548 lbs
Footprint 1,065 x 646 mm; 41.9 x 25.4“
Power supply 190 - 253 V AC ± 5 %, 50/60 Hz
Phase + neutral conductor + PE; the instrument
is equipped with a CEE 32A plug.
Instrument power consumption 4.0 kVA
Fuse protection of instrument 30 to 32 A (slow-blow)
Sound pressure level max. 55 dBA
Protection class IP 20 (IEC 60529)
Contamination class Contamination class II (laboratory environment)
Stability
(according to DIN EN 61010-1)
compliant if change of position <10°
The instrument is front-heavy.
Storage conditions 2 to 40 °C; 36 to 104 °F
Operating range (permissible ambient
temperature)
15 to 35 °C; 59 to 95°F
Recommended temperature range to
achieve the specified performance
18 to 25 °C; 64 to 77 °F, maximum temperature
gradient ∆T ≤ 3 °C/h. (5.4°F/h)
Humidity < 80% rel. (non-condensing)
Air quality No corrosive components and no high
concentrations of dust
Operation: Max. 3,000 m above mean sea level