Troubleshooting
— 70 SPECTRO ARCOS FHX3X — 08.02.2022
Malfunction Remedy
Asymmetrical plasma. Argon flushing from the OPI/SPI too high.
Reduce the flow rate.
Plasma chamber extraction set too high.
Reduce the extraction capacity.
Plasma flickers or
pulsates.
Moisture or deposits in the aerosol tube of the plasma torch.
Clean and dry the plasma torch or replace it.
Aerosol pulsates (deposits in nebulizer).
Clean the nebulizer. Replacing the pump tube.
Plasma too bright. Power is set too high.
Decrease the power.
Oxygen in sample introduction system.
Check for leaks and eliminate these.
Plasma extinguishes
when nebulizer gas flow
is switched on.
Sample introduction tube not connected to the nebulizer
Connect the tube and check fitting
Oxygen in sample introduction system.
Flush the system with argon before restarting system.
Leak in the sample introduction system.
Check for leaks and eliminate these.
No channel visible in the
plasma.
Nebulizer gas flow setting is too low
Optimize the nebulizer gas flow.
Nebulizer clogged.
Clean the nebulizer.
Leak in spray chamber outlet.
Check for leaks and eliminate these.
Sparkover at load coil. Humidity.
Switch off the generator and use a paper towel to dry the
load coil (if moisture is visible).
Dirt on the load coil
Switch off the generator and clean the load coil.