400.0 mV ±( 0.3% + 4 digits)
4.000, 40.00, 400.0 V ±( 0.5% + 3 digits)
600 V ±( 1.0% + 4 digits)
NMRR : >50 dB @ 50/60Hz
CMRR : >120 dB @ DC, 50/60 Hz, Rs=1 k
Input Impedance : 10 M, 30 pF nominal (1000 M for 400.0
mV range)
Transient protection : 6.5 kV (1.2/50 μs surge)

 
4.000, 40.00, 400.0 V ±( 1.5% + 5 digits)
600 V ±( 2.0% + 5 digits)
CMRR : >60dB @ DC to 60 Hz, Rs=1 k
Maximum Crest Factor: < 1.75 : 1 at full scale & < 3.5 : 1 at half scale
limited to fundamental and harmonics,
that fall within the meter specified AC
bandwidth for non-sinusoidal waveforms
Input Impedance : 10 M, 30 pF nominal
Transient protection : 6.5 kV (1.2/50μs surge)
ACD-10 Plus: Average Sensing
ACD-10 TRMS-PLUS: True RMS sensing - 5 % to 100 % of range