12
1) 2) 3)
40.00, 400.0, 600 A ±( 1.5% + 8 digits)
Overload Protections : ACA Clamp-on jaws : 600 A rms continuous
ACD-10 Plus: Average Sensing
ACD-10 TRMS-PLUS: True RMS sensing - 10 % to 100 % of range
1)
Max Induced error from adjacent current carrying conductor: 0.05 A
2)
Specified accuracy is from 1% to 100% of range and for measurements made at
the jaw center. When the conductor is not positioned at the jaw center, position
errors introduced are: Add 2% to specified accuracy for measurements made
BEYOND jaw marking lines (toward jaw opening)
3)
Add 8 digits to specified accuracy @ reading < 10% of range
400.0 Ω ±( 0.8% + 8 digits)
4.000, 40.00, 400.0 kΩ ±( 0.6% + 4 digits)
4.000 MΩ ±( 1.0% + 4 digits)
40.00 MΩ ±( 2.0% + 4 digits)
Open Circuit Voltage : 0.4 VDC typical
Transient protection : 6.5 kV (1.2/50μs surge)