EasyManua.ls Logo

Amprobe CO2-100 - SPECIFICATION; CO2 Specifications; Temperature and Humidity Specifications; Environmental and Power

Amprobe CO2-100
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
SPECIFICATION
CO2
Range
0~9999ppm
(5001~9999 out of accuracy scale range)
Resolution 1 ppm
Accuracy
±30ppm±5%rdg (0~5000)
(Not specified for out of scale)
Pressure
+1.6% reading per kPa deviation from
normal
Dependence Pressure, 100kPa
Temperature
Range -10.0~60.0°C (14~140°F)
Resolution 0.1°C /0.1°F
Accuracy ±0.6°C / ±0.9°F
Humidity
Range 0.0~95%
Resolution 0.1%
Accuracy ±3%(10~90% at 25°C) ; ±5%(others)
Operating
environment
0~50°C, 0~95%RH (avoid condensation)
Storage
environment
-20~60°C, 0~99%RH
(avoid condensation)
Power supply 4pcs AA batteries
- EMC: Conforms to EN61326-1. This product complies
with requirements of the following European Community Directives:
2014/30/EU (Electromagnetic Compatibility). However, electrical noise
or intense electromagnetic fields in the vicinity of the equipment
may disturb the measurement circuit. Measuring instruments will
also respond to unwanted signals that may be present within the
measurement circuit. Users should exercise care and take appropriate
precautions to avoid misleading results when making measurements
in the presence of electronic interference.

Related product manuals