ADE9000 Technical Reference Manual UG-1098
Rev. 0 | Page 49 of 86
Table 28. Service Type and VCONSEL, ICONSEL Setting Summary
Service Type
ADE9000
Ground
Reference
Figure for Reference
Number of
Voltage Sensors
Required
VCONSEL[2:0]
Number of
Current Sensors
Required
ICONSEL
4-Wire Wye
Neutral
Figure 62
3
000
3
0
4-Wire Wye Isolated Figure 63 3 000 3 0
3-Wire Delta Phase B Figure 64; Figure 62
with Phase B tied to
neutral
2 001;
VB = VA VC
2 0: IB has current sensor
1: IB = IA − IC
3-Wire Delta Isolated Figure 63 with
Phase B tied to
neutral
2 001;
VB = VA VC
2 0: IB has current sensor
1: IB = IA IC
3-Wire Delta Isolated Figure 66 2 100;
VA = VAVB;
VB = VA VC;
VC = VC − VB
2 0: IB has current sensor
1: IB = IA IC
4-Wire Delta Neutral Figure 62 (note that
VA and VB phasor
diagram follows
Figure 58)
3 000 3 0
4-Wire Wye,
Non-Blondel
Compliant
Neutral Figure 67 2 010;
VB = −VA VC
3 0
4-Wire Delta,
Non-Blondel
Compliant
Neutral Figure 68 2 011;
VB = −VA
2 0: IB has current sensor
1: IB = IA IC
3-Wire 1PH
Neutral
Not applicable
1 to 2
000
1 to 2
0
3-Wire
Network
Neutral Not applicable 2 000 2 0
Multiple 1PH
Circuits
Neutral Not applicable 3 000 3 0