Safety instructions
10 Edition 03/2013 SPECORD PLUS
2.6 Safety instructions - operation
2.6.1 General
The operator of the SPECORD PLUS must make sure before each commissioning that the
condition of the device is sound. This applies in particular after each modification or exten-
sion of the device or its repair.
Observe the following notes:
Free access to the device switch on the enclosure wall must be ensured during
operation.
Caution when handling glass components (cells or similar). Risk of broken glass
and therefore risk of injury!
Never look directly or indirectly, e.g. through a mirror, into the lamp beam! The de-
vice must therefore be switched off to replace the lamps! There is a risk of conjunc-
tivitis.
The ventilation equipment on the device must be in good working condition. Cov-
ered vents or ventilation slits etc. may cause the device to break down or may
cause damage to it.
Ensure that no liquid can enter the SPECORD PLUS. The device could be dam-
aged.
Do not place the SPECORD PLUS in the immediate vicinity of aggressive vapor,
e.g. strongly corrosive acid, lye or solvents! The vapor can corrode the terminals,
mechanical and optical components of the device.
Avoid contamination of the sample chamber with strongly corrosive substances or
highly volatile solvents! Use vapor-proof cells with sealing plug in this case. Do not
leave the samples longer than necessary in the sample chamber of the device after
the measurement!
2.6.2 Safety instructions - Protection against explosion and fire
The SPECORD PLUS may not be operated in an explosive environment.
The operating personnel must be familiar with the location of the fire-fighting equipment in
the operating room of the device.
2.6.3 Safety instructions - electrical equipment
Work on electrical components of the SPECORD may only be carried out by a qualified elec-
trician in accordance with the applicable electrical engineering rules. Life-threatening electri-
cal voltages occur at various locations within the system!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analytik Jena SPECORD 50 PLUS and is the answer not in the manual?

Analytik Jena SPECORD 50 PLUS Specifications

General IconGeneral
BrandAnalytik Jena
ModelSPECORD 50 PLUS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals