18
Section 5 Remote Control by GPIB
: PATTern: SEND: MODE <Character>
Function ............ The Pattern Generator delivery mode is set.
Parameter .......... <Character> = SINGLE
CONT
Unit ................... <Non term>
Restriction ......... This cannot be set during the data pattern delivery.
: PATTern: SEND: MODE?
Function ............ The Pattern Generator delivery mode is inquired.
Response ........... The Pattern Generator delivery mode = SINGLE
CONT
Restriction ......... None
: PATTern: RATE <Numeric> <Band rate term>
Function ............ The Pattern Generator delivery rate is set.
Parameter .......... <Numeric> = 1 bps to 99999 bps
Unit ................... <Band rate term: bps or kbps>
Restriction ......... None
: PATTern: RATE?
Function ............ The Pattern Generator delivery rate is inquired.
Response ........... The Pattern Generator delivery rate
Restriction ......... None
: PATTern: TYPE <Character>
Function ............ The Pattern Generator data output method is set.
Parameter .......... <Character> = 1bit
2bit
Unit ................... <Non term>
Restriction ......... This cannot be set in the Pattern Generator output ON status.
: PATTern: TYPE?
Function ............ The Pattern Generator data output method is inquired.
Response ........... The Pattern Generator data output method = 1bit
2bit
Restriction ......... None
: PATTern: SEND <Character>
Function ............ The Pattern Generator data pattern delivery execution/stop are performed.
Parameter .......... <Character> = RUN
STOP
Unit ................... <Non term>
Restriction ......... This cannot be executed in the Pattern Generator output OFF status.