Test Parameter Variables
MT8852B Programming Manual PN: 13000-00110 Rev. V 12-23
PTXLEV (MT8852B and MT8852B-042 only)
This parameter is used to define whether the EDR Relative Transmit power test (ERPCFG)
should be carried out at minimum and/or maximum power levels.
PWRDELAY
This parameter is used to set or read the delay required for the EUT to change the Tx power
as requested before measurements are made. Bluetooth devices that support power control
should have this time specified in the ‘Implementation Extra Information for Testing’ (IXIT)
document.
Set command format ERPCFG<ws><script number><,>PTXLEV<,><pow level>
<script number> 3 to 10
<pow level> MIN | MAX | MINMAX
Example To set the ERPCFG to perform the test at both minimum and
maximum power for script 7 the command will be:
ERPCFG 7,PTXLEV,MINMAX
Query command format ERPCFG?<ws><script number><,>PTXLEV
<script number> 1 to 10
Response The response string returned for the query is in the identical
format as the configuration command string.
Example ERPCFG? 4,PTXLEV
Response If the power level was set to maximum the command will be:
ERPCFG 4,PTXLEV,MAX
Set command format PCCFG<ws><script number><,><PWRDELAY<,><value>
<script number> 3 to 10
<value> 100 Milliseconds to 100 seconds in seconds
(1 sec default)
Example To set the delay to 1 second in script 4, the command would be:
PCCFG 4,PWRDELAY,1
Query command format PCCFG?<ws><script number><,>PWRDELAY
<script number> 1 to 10
Response The response is returned in the form of the command to set that
state.
Example PCCFG? 7,PWRDELAY
Response If script 7 delay is set to 5 seconds, the response would be:
PCCFG 7,PWRDELAY,5.0e+000