Appendix A — Test Records A-3 Test Records for Spectrum Analyzer Verification
S412E MM PN: 10580-00324 Rev. N A-3
S412E Firmware Rev: ______________ Operator: ____________________ Date: _____________
Serial Number: _______________ Options: ___________________________________________________
A-3 Test Records for Spectrum Analyzer Verification
Table A-3. Spectrum Analyzer Residual Spurious Response with Preamp Off
Start Freq Stop Freq RBW VBW Specification Measured Values
10 MHz 1.6 GHz 3 kHz 1 kHz <
–90 dBm dBm
(PERFORM THE FOLLOWING TESTS IF OPTION 6 IS INSTALLED)
1.6GHz 6.0GHz 3kHz 1kHz <
–90 dBm dBm
Table A-4. Spectrum Analyzer Residual Spurious Response with Preamp On
Start Freq Stop Freq Specification Measured Values
10 MHz 1.6 GHz <
–90 dBm dBm
(PERFORM THE FOLLOWING TESTS IF OPTION 6 IS INSTALLED)
1.6GHz 6GHz <
–90 dBm dBm
Table A-5. Spectrum Analyzer DANL with Preamp Off
Start Freq Stop Freq RBW VBW Specification
Measured Values
at 100 kHz RBW
Calculated Values
for 10 Hz RBW
10 MHz 1.6 GHz 100 kHz 1 kHz <
–131 dBm dBm dBm
(PERFORM THE FOLLOWING TESTS IF OPTION 6 IS INSTALLED)
1.6 GHz 2.4 GHz 100 kHz 1 kHz <
–131 dBm dBm dBm
2.4 GHz 4.0 GHz 100 kHz 1 kHz <
–127 dBm dBm dBm
4.0 GHz 5.0 GHz 100 kHz 1 kHz <
–124 dBm dBm dBm
5.0 GHz 6.0 GHz 100 kHz 1 kHz <
–116 dBm dBm dBm