Table – 9 Temperature compensation coefficient of special solutions
Solution
Temperature compensation coefficient
NaCl solution
2.12%/°C
5% NaOH solution
1.72%/°C
Dilute ammonia solution
1.88%/°C
10% hydrochloric acid solution
1.32%/°C
5% sulfuric acid solution
0.96%/°C
6.2.5 Precaution for calibration solution’s contamination
Conductivity standard solution has no buffer. Please avoid being contaminated during usage. Before
submerging the electrode in standard solution, please rinse the electrode in distilled water and dry it with
clean cloth or tissue paper. Please do not use the same cup of conductivity standard solution repeatedly,
especially for standard solution of low concentration (84μS/cm). The contaminated standard solution will
affect the accuracy.
6.3 Conductivity Calibration
6.3.1. Rinse the electrode in distilled water, allow it to dry, wash with a little of
standard solution and dip it in standard solution. Stir the solution briefly and allow
it to stay in the solution until a stable reading is reached.
6.3.2. Press key to enter the calibration mode.
The meter’s display will show blinking CALat the top right, and scanning and
locking process of calibration solution at the bottom right. When the meter
locks1413 mS, stable icon will display on LCD. Press key to complete
calibration.
The meter will return to measuring mode and is displayed on bottom
left of the LCD screen. See Figure-11.
6.3.3. Notes:
(a) Pressing key before stable icon appeared on LCD screen,
Error indication icon Er 2 will be shown. See Table -11.
(b) Press key before confirmation to exit calibration mode (calibration will not be
completed).
6.3.4. For multi-point calibrations, please repeat the steps in 6.3.1 to 6.3.2 until all calibrations are finished.
The meter can be calibrated in the same calibration solution repeatedly until the reading is stable and
repeatable.
Figure-6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Apera Instruments EC400 and is the answer not in the manual?

Apera Instruments EC400 Specifications

General IconGeneral
BrandApera Instruments
ModelEC400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals