EasyManuals Logo

Apera Instruments PC60 User Manual

Apera Instruments PC60
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #4 background imageLoading...
Page #4 background image
!
4!
3. Complete Kit
4. Preparation Before Use
If it is first-time use or the tester hasn’t been used for a long time, pour some 3M KCL solution in
the probe cap (about 1/5 of the probe cap) and soak the probe for 15-30 minutes.
When not in use, we recommend store the pH probe in the storage 3M KCL solution in the probe
cap to keep the sensor’s accuracy. But even if stored dry, it won’t do any permanent damage to
the sensor. It will only temporarily cause the probe to lose its sensitivity, which can always be
restored by soaking in the storage solution.
The!storage!solution! is! 3M!KCL! (potassium! chloride).!One! bottle! of! 10m L! storage! solution!com es! with! the!
tester!kit.!If!the!soaking!solution !w as!co nta m ina ted ,!please !replac e!w ith!new!ones!timely.!
*!DO!NOT!use!any!other!brand’s!storage!solutions!because!different!chemicals!may!be!used!and!
potential!permanent!damage!could!be!caused!to!the!meter.!
Diagram!-!2!

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Apera Instruments PC60 and is the answer not in the manual?

Apera Instruments PC60 Specifications

General IconGeneral
BrandApera Instruments
ModelPC60
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals