SYST:ENER:IMP:MOD?
Request for the state of the device impulse output. Reply is number: 0 … state is FIXED, 1 … state is
ACTIVE POWER, 2 … state is REACTIVE POWER, 3 state is APPARENT POWER
SYST:ENER:IMP:MOD {<mode>}
Setting the state of device’s impulse output. Parameter is number: 0 state is FIXED, 1 state is
ACTIVE POWER, 2 … state is REACTIVE POWER, 3 … state is APPARENT POWER
SYST:ENER:IMP:AUX?
Request for the state of AUX output. The same parameter as described in Measurement Settings.
Answers are: 0 = OFF, 1 = ON.
SYST:ENER:IMP:AUX {<auxiliary>}
Setting of the state of AUX output. The same parameter as described in Measurement Settings.
Parameter can be 0 = OFF, 1 = ON.
SYST:ENER:DOS:INIT
Initialization of dosage function.
SYST:ENER:DOS:STAR
Start (or initialization) of dosage function.
SYST:ENER:DOS:STOP
Stop (or cancel) of dosage function.
SYST:ENER:DOS?
Request for general status of dosage function. If dosage function is active reply is: percentage, done
energy or time and full energy or time.
SYST:ENER:DOS:STAT?
Request for status of dosage function. Reply is 0 (dosage not active), 1 (dosage initialized), 2 (dosage
started and running) or 3 (dosage stopped).
SYST:ENER:DOS:STAT:PERC?
Request for done percentage of dosage.
SYST:ENER:DOS:STAT:VAL?
Request for done value (time or energy amount) of dosage.
SYST:ENER:DOS:STAT:TIM?
Request for time of “time dosage” in seconds.
SYST:ENER:DOS:STAT:ENER?
Request for energy of “energy dosage” in Ws resp. VAs.
SYST:ENER:DOS:STAT:ENER:K?
Request for done energy of “energy dosage” in kWh resp. kVAh.
SYST:ENER:DOS:MOD {<mode>}
Setting of Dosage Mode. Parameter is number (0 - 5) or code (TIME, ACTIVE …) of Dosage Mode.
SYST:ENER:DOS:MOD?
Request for state of Dosage Mode. Answer is number (0 - 5) or code (TIME, ACTIVE …).

Other manuals for Applied Precision WS 2120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Precision WS 2120 and is the answer not in the manual?

Applied Precision WS 2120 Specifications

General IconGeneral
BrandApplied Precision
ModelWS 2120
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals