EasyManuals Logo

Applied Precision WS 2120 User Manual

Applied Precision WS 2120
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #82 background imageLoading...
Page #82 background image
Working Standard
User’s Guide
WS-UGFW619-EN www.appliedp.com
82
SYST:ENER:DOS:MAXD:SUBP {<subperiod>}
Setting the value of Maximum Demand Test’s sub-period parameter.
SYST:ENER:DOS:DUT:MAXD:K {<register>}
Setting the register value of Maximum Demand Test for DUT (Device Under Test).
SYST:ENER:DOS:RES:MAXD?
Request for error of Maximum Demand Test after finishing this test.
SYST:ENER:DOS:STAT:MAXD:K?
Request for state of Maximum Demand Test.
SYST:ENER:DOS:STAT:MAXD:K:TEMP?
Request for the recently integrated value of power (Maximum Demand Test).
SYST:ENER:DOS:STAT:MAXD:WIN:BEG?
Request for the beginning timestamp of Maximum Demand Test being in progress.
SYST:ENER:DOS:STAT:MAXD:WIN:END?
Request for the ending timestamp of Maximum Demand Test being in progress.
SYST:DOS?
Request for state of Dosage mode. Answer is 0 (=disabled) or 1 (=enabled).
SYST:DOS:ENAB {enable(0|1) }
Setting of state of Dosage mode. Parameter enable can be 0 (=disable) or 1 (=enable).
SYST:DOS:FALL?
Request for Dosage fall method. Answer is 0 (=soft generate) or 1 (=immediate).
SYST:DOS:FALL { method(0|1) }
Setting of Dosage fall method. Parameter can be 0 (=soft generate) or 1 (=immediate).
SYST:DOS:RISe?
Request for Dosage rise method. Answer is 0 (=soft generate) or 1 (=immediate).
SYST:DOS:RIS { method(0|1) }
Setting of Dosage rise method. Parameter can be 0 (=soft generate) or 1 (=immediate).
SYST:DOS:STOP:VOLT { state (0|1) }
Setting of voltage Stopping feature after Dosage completion. Parameter can be 0 (disabled) or 1
(enabled).
SYST:DOS:STOP:VOLT ?
Request for state of “voltage Stopping feature after Dosage completion”. Answer can be 0 (disabled) or
1 (enabled).
SYST:DOS:STOP:CURR { state (0|1) }
Setting of current Stopping feature after Dosage completion. Parameter can be 0 (disabled) or 1
(enabled).
SYST:DOS:STOP:CURR ?
Request for state of “current Stopping feature after Dosage completion”. Answer can be 0 (disabled) or
1 (enabled).

Other manuals for Applied Precision WS 2120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Precision WS 2120 and is the answer not in the manual?

Applied Precision WS 2120 Specifications

General IconGeneral
BrandApplied Precision
ModelWS 2120
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals