Water Activity Meter
Operator’s Manual
For Series 4, 4TE, 4TEV, DUO, and TEV DUO
Decagon Devices, Inc.
Version: August 2, 2016 — 09:05:15

Table of Contents

Questions and Answers

  • D
    Dakota DavenportAug 27, 2025
    How to fix AquaLab 4TEV that does not turn on?
    • J
      Jacob HansenAug 27, 2025
      If your AquaLab Measuring Instruments does not turn on, check to make sure your power cord is securely attached to the back of the instrument and plugged into the power outlet. A power surge may have caused a fuse to blow, follow instructions to change the fuses. If the fuse blows again, a failed component may be causing the problem.
  • S
    Selena FarmerAug 27, 2025
    Why is moisture content not correct on AquaLab 4TEV Measuring Instruments?
    • M
      middletonpaulAug 27, 2025
      If the moisture content displayed on AquaLab Measuring Instruments is not correct, it may be because the model selected is not correct for the product being tested. Toggle through the available models to find a more appropriate one. If the model is correct but still not giving correct moisture content values, it may be necessary to generate a new model for the product or update an existing model.
  • L
    Laura FlynnAug 25, 2025
    What to do if AquaLab Measuring Instruments displays aw below instrument detection limits?
    • J
      Joshua WattsAug 26, 2025
      If your AquaLab Measuring Instruments displays aw below instrument detection limits, it means the sample is too dry for the instrument to read accurately. If your sample has a water activity that is less than the detection limits of the instrument, this message appears. You can also try cleaning the mirror and chamber and measuring the sample again.
  • K
    kevinhouseAug 23, 2025
    How to fix 'Contaminated mirror' message on AquaLab 4TEV Measuring Instruments?
    • M
      Madeline BarrAug 25, 2025
      If the AquaLab Measuring Instruments displays the message “Contaminated mirror,” it means the mirror used for dew point measurements requires cleaning. Follow the cleaning instructions.

Summary

Introduction to AquaLab Water Activity Meter

General Safety Information

Important guidelines and warnings for safe operation of the AquaLab instrument.

Getting Started with AquaLab

Preparing AquaLab for Operation

Steps to set up and prepare the AquaLab before first use.

AquaLab Menus and Navigation

Cleaning and Maintenance Procedures

Cleaning Procedure

Detailed step-by-step instructions for cleaning the AquaLab instrument.

Verification of Calibration

Procedures to check and ensure the accuracy of the AquaLab's calibration.

Verification and Calibration

Water Activity Verification

Daily checks to guarantee optimal performance and accuracy against standards.

Sample Preparation Guidelines

Taking a Water Activity Reading

DUO Operation (Optional)

Computer Interface and Software

Troubleshooting Common Issues

Instrument Power and Startup Issues

Solutions for problems related to powering on the AquaLab.

Slow or Inconsistent Readings

Troubleshooting steps for slow measurements or unstable results.

Sensor and Calibration Problems

Diagnosing and resolving issues with sensors or calibration accuracy.

Support and Repair Information

Appendix A: Preparing Salt Solutions

Appendix B: Temperature Correction of Verification Standards

Appendix C: AquaLab Verification Standards Application Note

AquaLab 4TEV Series Specifications

General IconGeneral
Accuracy±0.003 aw
Temperature Control15 - 40 °C
Temperature Accuracy±0.1 °C
Power Requirements110 - 220 VAC, 50/60 Hz
DisplayColor touch screen
Water Activity Range0.030 - 1.000 aw
Measurement TimeTypically 5 minutes
CommunicationUSB
Measurement Range0.030 - 1.000 aw
ConnectivityUSB
Operating Humidity0 - 90% RH non-condensing