Ardo medical AG 19
Operating Instructions
Master and Senator
Subject to modifications
9 .
T
E C H N I C A L
I
N F O R M A T I O N
9.1 Technical Data
Electrical data for 230VAC devices
Mains supply voltage 230VAC
Mains frequency 50Hz
Power consumption
Master 140W / 0.6A
Senator 115W / 0.45A
Device fuse L/N T1.25AH / AC250V
Electrical data for 115VAC devices
Mains supply voltage 115VAC
Mains frequency 60Hz
Power consumption
Master 140W / 1.2A
Senator 115W / 1.0A
Device fuse L/N T1.60AH / AC250V
Performance data
Suction category high vacuum / high flow
Suction power max.
Master 50l/min ±15%
Senator 30l/min ±15%
Vacuum meter 0 to -100kPa / 0 to -700mmHg / ±2.5%
Vacuum regulator mechanically, continuously 0 to maximum
Vacuum height max. at
2'000m a.s.l.
-72kPa / -540mmHg / ±5%
1'000m a.s.l.
-84kPa / -630mmHg / ±5%
500m a.s.l.
-89kPa / -668mmHg / ±5%
0m a.s.l.
-95kPa / -713mmHg / ±5%
Classification
Protection class Class I with protective earth ground
Degree of protection BF
Classification according MDD 93/42/EEC IIa
Conformity for 230VAC devices
Concerning Medical Device Directive
MDD 93/42/EEC 0123
EN 60601-1 / EN60601-1-2 / EN60601-1-6 /
EN ISO 10079-1 / EN ISO 14971
Conformity for 115VAC devices
With respect to electric shock, fire and mechanical
hazards only in accordance with UL 2601-1 / 52YA.
With respect to electric shock, fire and mechanical
and other specified hazards only in accordance with
CAN/CSA C22.2 No.601.1, (CAN/CSA 601.2XX, if
applicable) Medical equipment certified for Canada
52YA E218915.
E218915
UL2601-1
52YA