16
pIf the battery charge is too low, the NOMAD Pro 2 will
not emit an X-ray and the Recharge Handset alarm will
display.
qWhen a battery is depleted to the single bar level
it is recommended the handset be recharged. Remove
the handset from the charging cradle once it is fully
charged.
rReplace a low charge handset with the secondary
handset (which should be charged in advance). It is
recommended that you keep one handset fully charged at
all times to ensure continuous operation.
3.4 Optional Checks
The NOMAD Pro 2 is factory calibrated and tested prior
to release (see your Certificate of Conformance). A self-
diagnostic is completed each time an exposure is taken.
However, the optional checks listed below may be performed periodically as desired. Some
locations may require initial and periodic checks as a condition of use.
The NOMAD Pro 2 has an X-ray disable feature that allows the X-rays to be disabled for
training purposes. Test firing this X-ray unit may otherwise be dangerous to the testing
technician or bystanders. See
Section 4.2 for more information.
Power On/Off: Attach a fully charged handset to the NOMAD Pro 2 device. Press the
Power
button and verify that the device correctly powers on and off.
Increase/Decrease Exposure Time: With the device on, select the patient size,
the image receptor type, and the tooth type. The exposure time corresponding to
the selections will display on-screen. Press the Increase
and the Decrease
buttons, and verify that the exposure time changes appropriately. When the
Increase
or Decrease buttons are pressed and held, the rate at which the
displayed time setting changes on screen accelerates.
Triggering: With the device on and an exposure time selected, press and release
the trigger once. This will ready the device, and the display will indicate the Ready
state. To end the Ready state, press any button on the user interface panel or wait
approximately 15 seconds for the timeout.
Automatic Shut-down: With the device on, allow approximately three minutes of
inactivity to pass for the system to automatically shut down.
The NOMAD Pro 2 should not be operated if it has been dropped or if performance
degrades; it should be returned to Aribex or an authorized service center for an evaluation.
(See
Section 6.1 Alarms and Alerts for more information.)
Recharge Handset Alarm