OPERATOR MANUAL
MANUFACTURED BY ARIBEX
Handheld X-ray System
for Intraoral Radiographic Imaging

Questions and Answers

  • K
    Kari GraySep 23, 2025
    Why ARIBEX Nomad Pro 2 indicator lights illuminate orange on charging cradle?
    • J
      jeremysimpsonSep 23, 2025
      If the indicator lights illuminate orange when first placing the handset of your ARIBEX Dental equipment on the charging cradle, the male connectors on the charging cradle may be damaged. In this case, the NOMAD Pro 2 will require authorized service.
  • W
    Willie PerryAug 23, 2025
    How to improve contrast in ARIBEX Nomad Pro 2 X-ray images?
    • A
      Andrew ArcherAug 25, 2025
      An X-ray image might lack sufficient contrast due to underexposure (too light) or overexposure (too dark). If underexposure is the cause, increase the exposure time. If it's overexposure, decrease the exposure time and check the film expiration date, as old film can produce dark, grainy images. Also, if you are using film-based imaging, ensure your chemical developer is fresh and at the proper temperature.
  • A
    alexandriarichardsAug 25, 2025
    What to do if the 'Incomplete Exposure' alarm appears on my ARIBEX Dental equipment?
    • C
      Carla RushAug 26, 2025
      The 'Incomplete Exposure' alarm on your ARIBEX Dental equipment appears when the trigger is released before the timed exposure completes. To resolve this, ensure you depress the trigger for the full duration of the timed exposure. The alarm will automatically clear within 15 seconds or by pressing any button on the user interface panel.
  • E
    Eric MartinezAug 27, 2025
    What does the 'System Failure' alarm mean on my ARIBEX Nomad Pro 2?
    • G
      gabriel65Aug 27, 2025
      The 'System Failure' alarm on your ARIBEX Dental equipment indicates that self-diagnostics detected a potential issue during X-ray firing. You can clear this warning alarm by powering the device off and then on again. If the alarm reappears, the NOMAD Pro 2 will require authorized service.
  • T
    Tyler HansenAug 27, 2025
    What to do if the 'Handset Failure' alarm appears on my ARIBEX Dental equipment?
    • C
      christinawalkerAug 27, 2025
      The 'Handset Failure' alarm on your ARIBEX Dental equipment indicates that the handset is not communicating with the device. First, ensure the handset is securely attached. If the issue persists, the protective fuse in the handset may be blown, or the batteries may be impaired. Replace the handset with a newly charged one. Check the charging cradle: if the charging bars illuminate solid red or not at all, the handset will require authorized service.
  • P
    Peter NguyenSep 17, 2025
    What does 'incomplete' mean on ARIBEX Dental equipment?
    • B
      Brianna YoungSep 17, 2025
      If the 'INCOMPLETE' message is displayed on the screen of your ARIBEX Dental equipment, be sure to depress the trigger for the duration of the timed exposure. This condition automatically clears within 15 seconds or by pressing any button on the user interface panel.
  • S
    smithdanielleSep 14, 2025
    How to troubleshoot ARIBEX Nomad Pro 2 Dental equipment control panel with no power?
    • S
      Samantha BakerSep 14, 2025
      If there is no power to the ARIBEX Dental equipment control panel, ensure the handset is securely attached. If the display does not illuminate after pressing 'Power' several times, the handset is likely the problem. You can also replace it with a newly charged handset. Use the charging cradle to determine if the first handset battery can be recharged.
  • L
    lance52Sep 15, 2025
    Why does ARIBEX Nomad Pro 2 Dental equipment ready indicator terminate early?
    • N
      Nancy AllenSep 15, 2025
      If the READY indicator on your ARIBEX Dental equipment terminates before an exposure starts, double-check the exposure time setting and re-enable when operator and patient are again properly situated. The NOMAD Pro 2 READY condition expires if the X-ray exposure is not initiated within 15 seconds of the start of the READY condition.
  • W
    william51Sep 6, 2025
    How to fix insufficient contrast in ARIBEX Nomad Pro 2 Dental equipment X-ray?
    • N
      Nicole SandersSep 6, 2025
      If the image from your ARIBEX Dental equipment X-ray exposure lacks sufficient contrast due to underexposure (image is too light), increase the exposure time setting. You can also refer to the instructions for 'Incomplete Exposure'.
  • D
    Deborah SampsonSep 3, 2025
    What to do if there is no power to my ARIBEX Dental equipment NOMAD Pro 2 control panel?
    • L
      lewisbethanySep 3, 2025
      If there is no power to the NOMAD Pro 2 control panel of your ARIBEX Dental equipment, and pressing the power button several times doesn't illuminate the display, the attached handset is likely the problem. Ensure the handset is securely attached. If all handsets appear fully charged, the NOMAD Pro 2 will require authorized service.

Summary

Safety Precautions

Radiation Safety

Guidelines for safe operation, operator protection, and minimizing radiation exposure.

Studies and Data on Leakage and Scatter

Presents data on radiation leakage and scatter, comparing NOMAD Pro 2 exposure levels to limits.

Setup and Power Check

The Backscatter Shield

Describes the function and adjustment of the backscatter shield for operator protection.

Operation

Ensuring the Right Exposure Time is Set

How to select and set appropriate exposure times based on patient and imaging parameters.

Initiating and Completing an X-ray Exposure

Detailed process for starting and finishing an X-ray exposure using the trigger.

Exposure Techniques

Guidance on positioning the device and patient for optimal intra-oral X-ray imaging.

Ensuring Image Quality

Features contributing to image quality and tips for achieving optimal results.

Technique Factors Settings and Adjustments

Provides pre-set technique factors and guidance for adjusting them based on patient needs.

NOMAD Pro 2 Care and Upkeep

Troubleshooting

A guide to diagnosing and resolving common operational problems and error messages.

ARIBEX Nomad Pro 2 Specifications

General IconGeneral
TypeHandheld X-ray System
Tube Voltage60 kV
Tube Current2.5 mA
Focal Spot0.4 mm
Total Filtration1.5 mm Al equivalent
DisplayLCD
UsageIntraoral Radiography
X-ray GeneratorHigh Frequency
X-ray SourceX-ray tube
Weight2.5 kg (5.5 lbs)
Battery TypeLithium Polymer
Image SensorNot applicable (external sensor required)
Battery LifeUp to 400 exposures per charge