42
Test Type
In Accordance with
Document Document Title
Power Frequency Magnetic
Field Immunity
IEC 61000-4-8:1993,
A1:2000
Electromagnetic Compatibility, Testing
and Measurement Techniques for Power
Frequency Magnetic Field, Immunity Test
NOTE: The tests documented in the table above are the only tests required for this product
as it is a battery operated device and the AC Adapter (CP-0062) is certified. IEC 61000-3-2,
3-3, 4-4, 4-5, 4-6, 4-11, and EN 55014 1:2006 are not applicable.
Emissions Test Summary
Specification Frequency Range Compliance Status
EN 55011:2007, Group 1,
Class “B” Conducted Emissions
0.15MHz – 30.00MHz PASS
EN 55011:2007, Group 1,
Class “B” Radiated Emissions
30.0MHz – 2500MHz PASS
Immunity Test Summary
Specification
Minimum
Test Level
Required per EN
60601-1-2
Test Level
Completed
Compliance
Status
IEC 61000-4-2:1995, A1:1998,
A2:2000 – Electrostatic
Discharge Immunity
Air discharge up to
±8kV; contact
discharge up to
±6kV
Air discharge up to
±8kV; contact
discharge up to
±6kV
PASS
IEC 61000-4-3:2006 RF
Radiated Fields Immunity
Radiation field
strength of
3V/m from 80 –
6000MHz (80% AM
@ 1kHz)
Radiation field
strength of
3V/m from 80 –
6000MHz (80% AM
@ 1kHz)
PASS
IEC 61000-4-8:1993, A1:2000
Power Frequency Magnetic Field
Immunity
Helmholtz coil at
50Hz and 60Hz, to
3 amps (rms) per
Helmholtz coil at
50Hz and 60Hz, to
3 amps (rms) per
meter
PASS