11
2.6 Display options
A service technician can set the information to be shown in the
display.
2.6.1. Displayed while a program is running
The display can be set to show one of the following alternatives
while a program is running:
Temperature
A
0
value
A
0
value / Temperature
Remaining Time
Temperature / Remaining time
A
0
value / Remaining time
A
0
value / Remaining time / Temperature
If an alternative with more than one parameter is selected, the
parameters will be displayed alternately. The A
0
value always
remains at 0 until the thermal disinfection begins. When the
temperature reaches 85°C, the A
0
value will increase in line with
the temperature and time.
2.6.2. Display after program end
The display can be set to show one of the following alternatives
once a program has nished:
Max. temperature
A
0
value
A
0
value / Max temperature
If an alternative with more than one parameter is selected, the
parameters will be displayed alternately.