17
4.3.2. Service table
The machine is connected to the electricity supply
and some components are live.
Maintenance may only be performed by authorized
service technicians.
NOTE! The service table is only shown for information
purposes.
The service table below shows recommended maintenance
intervals.
In addition to system testing, we recommend that appropriate
cleaning tests and temperature validation are performed according
to EN ISO 15883.
Measure Every year /
10,000 cycles
Every other
year /
20,000 cycles
Time*
(minutes)
1 General
1.1 Check the cabling and connection points. 10
1.2 Check that the panel decal is rmly a󰀩xed,
is undamaged and legible.
2
1.3 Check that an goods placement sign and
quick help instructions have been put up.
1
1.4 Check error codes and the number of
processes.
2
2 The chamber
2.1 Check that the rotating nozzles are moving
freely. Clean as required.
20
2.2 Check that the xed nozzles are not
clogged with grime and deposits. Clean as
required.
20
2.3 Check for leakage in the nozzle
attachments and the hose connections to
the chamber.
-
2.4 Check for proper operation of the
temperature sensor.
10
2.5 Check that the door seal and chamber seal
are tight. Clean or replace as necessary.
10

Other manuals for Arjo Typhoon