Page 54 · 5400 Series TECSource User’s Manual
Error
Code
Description Cause
E-124 Data mismatch
The message contained data that did not
match the expected format (USB/Serial
only).
E-126
Too few or too many
elements
The command requires more or less than
the number of parameters actually
supplied.
E-127 Change not allowed
An attempt was made to change a
parameter that cannot be changed, or is
currently read-only.
E-201 Data out of range
The message attempted to set a value that
was outside the allowable range
(USB/Serial only).
E-202 Invalid data type
When trying to parse the message, the data
was in an invalid format (USB/Serial only).
E-204 Suffix not valid
An invalid number base suffix (radix) was
encountered when parsing a number
(USB/Serial only).
E-217
Configuration Recall
failed
An attempt recall a configuration failed.
This can be caused if no configuration
exists in the selected slot, the slot number
is out of range, or if the configuration is
corrupted.
E-218
Configuration Save
failed
An attempt to save a configuration failed.
This can be caused if the slot number is out
of range, or the configuration memory is
corrupted.
E-220 Script Save Failed
An attempt to save a script failed. This can
be caused if the script number is out of
range, or the script memory is corrupted.
E-221 Cannot embed script
A script was executed that contained a
reference to another script.
E-222 Cannot execute script
An attempt to execute a script failed. This
can be caused if the script number is out of
range, no script exists for the selected
index, or the script memory is corrupted.
E-303 Input buffer overrun
The command input buffer overran, which
can be caused by excessively long
command strings or improperly terminated
commands.
E-402
Sensor open, output
turned off
A sensor open circuit was detected and the
output was turned off.
E-403
Module open, output
turned off
A Peltier module open circuit was detected
and the output was turned off.