Cleaning, disinfection and sterilization
72 1590 99 002_1 - 21. February 2011
Improper use or maintenance of the laser system may void the warranty granted by Asclepion Laser
Technologies GmbH. Before starting any repair or if you have questions that have not been
answered by this manual, contact your service partner (see chapter 10 Fehler! Verweisquelle konnte
nicht gefunden werden.).
9.2 Fault messages and fault removal
The MeDioStar is provided with alarm systems. If they are released, the system will switch to the
safe mode and stop the emission of laser radiation. Please try to remove the fault yourself by
following the instructions given in the table below. If the error persists, contact Asclepion Laser
Technologies GmbH or a service partner expressly authorized by Asclepion Laser Technologies
GmbH.
Before you contact the Technical Service of Asclepion Laser Technologies GmbH, please make a
note of the fault message and the error code displayed.
Fault message Explanation / Corrective action
Communication e
r
r
o
r
(all 50 XXX errors)
The communication is interrupted.
Start the device anew. Operation can be continued. Please contact the
Technical Service if the error occurs again.
Temperature error
(all 150 XXX errors
)
The operating temperature is too low or too high.
Check if the device is sufficiently ventilated. Start the device anew.
Operation can be continued. Please contact the Technical Service if t
he
error occurs again.
Foot switch
(400 XXX errors)
Please release the foot switch.
(do not activate the foot switch too early, wait until the display shows
READY with yellow warning sign)
Door interlock
(600 601 errors)
The door interlock contacts are o
pen or the door interlock plug is not
connected. If the door has been co
nnected to the door interlock, door
open.
Connect the door interlock contacts correctly or activate the door interlock
switch. If the door has been connected to the door interlock, clo
se the
door.
Handpiece error
(all 110X errors)
Check the handpiece connection.
Start the device anew. Please contact the Technical Service if the error
occurs again.