EasyManuals Logo

ATEQ D6 Series Quick Start Guide

ATEQ D6 Series
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
QSG_D6 CLASS_439.00_EN_01 /_2020-05-04
7/38
FLOW RATE TEST
The ATEQ D6 series can do direct or indirect measurements. In both cases, the flow meter
can also work in vacuum (optional).
Measurement principle
When the fluid (gas) enters the device
1
, it moves through a calibrated flow tube
2
which
causes a drop in pressure. The pressure drop is measured by a differential pressure sensor
3
. The pressure of the part under test
5
is measured by the sensor
4
.
1 Device
2 Calibrated flow tube
3 Differential pressure sensor
4 Pressure sensor
5 Part under test
Direct measurement
Once the part is filled, the device measures the flow rate through the part.
The test pressure is applied to the device 1 and then to the input of the part under test 2.
1 Device
2 Part under test