SYMPTOM PROBABLE CAUSE SOLUTION
Scaledoesnot
turnon
AC Adapter is not connected securely
Re-plug the AC adapter or rotate the plug
to securely connect it to the scale
Low baery Replace the baeries
Indicator is damaged
Replace with a new indicator and perform
calibraon
The cable from the plaorm to indicator is
not correctly connected, is disconnected,
or has short circuited
Check connecons. If damaged return the
scale for repair
Indicator is damaged
Replace with a new indicator and perform
calibraon
Load cell cable is broken Return the scale for repair
Load cell is damaged Return the scale for repair
Weight reading exceeds Power On Zero limit
1. Ensure scale plaorm is empty
2. Perform zero calibraon.
3. Reduce the weight on the plaorm
unl the indicaon is within the
key zero range
Indicaon is out of key zero range
Weight reading below Power On Zero limit
1. Install plaorm on the scale
2. Check for objects stuck between the
load cell and the scale base, remove
if present
3. Perform zero calibraon
Weight reading exceeds overload limit
Reduce load on scale unl the weight
value is displayed
Weight value can’t be displayed in the
current unit of measure because it
exceeds 6 digits
Use a more appropriate unit of measure
Weight reading below Under Load Limit
1. Install plaorm on scale
2. Perform zero calibraon