IFU-00367 Rev A
23
9. Operating & Storage Conditions
Operating Conditions
The Avive AEDä is intended for use in professional or home healthcare environments, including public and
private locations. It should only be used in the following environmental conditions:
Temperature: 32˚F 122˚F (0˚C 50˚C)
Battery Charging: 32˚F – 113˚F (0˚C – 45˚C)
Humidity: 5 95% non-condensing
Altitude: -280 ft. to 15,000 ft. (-85 m to 4572 m)
Water & Ingress: The Avive AED is rated to IP54 (with Aviveä Pad Cartridge installed) - dust protected, splash
and spray proof.
IMPORTANT: Do not expose the Avive AED to any liquids or dust when a Pad Cartridge is not installed.
Storage Conditions
It is important that the Avive AED be stored within the following environmental ranges. If the device is stored
outside of these ranges, then the device may be damaged and no longer functional for emergency use.
Long-term Storage Temperature: 41˚F to 104˚F (5˚C to 40˚C)
Short-term Storage Temperature (Up to 14 Days): -4˚F to 140˚F (-20˚C to 60˚C)
Humidity: 5 – 95% non-condensing
Altitude: -280 ft. to 15,000 ft. (-85 m to 4572 m)
IMPORTANT: If the device has been stored outside of 0˚C-50˚C, then it should be returned to ambient
temperature for at least 24 hours.
STORAGE TIP:
To keep your device healthy, looking good, and maximize lifetime, store it:
• In a clean and dry place
• At room temperature
• Out of direct sunlight