IFU-00367 Rev A
11. Technical Data
Physical Specifications (with Avive
ä
Pad Cartridge Installed)
Size: 143mm x 160mm x 52mm (5.63in x 6.30in x 2.05in)
Weight: 0.73 kg (1.6 lbs.)
Battery
Battery Type: 3.635V, 3500mAH, Lithium Ion, Rechargeable, not user replaceable
Time to Full Battery Charge: <5 hours
Capacity [Battery Fully Charged]: ≥75 shocks at 150J or 2.5 hours of continuous operation
Low Battery Indicator: < 10 150J Shocks, < 20 minutes continuous operation.
Maximum Battery Standby Time until Low Battery : ≥ 8.5 Months
Pad Cartridge Specifications
Electrode Type: Single-use combined ECG / defibrillation multifunction electrode.
Type BF Applied Part
Electrode Placement: Adult: Anterior-Lateral
Pediatric: Anterior-posterior
Electrode Active Area: >75 cm
2
Electrode Cable Length: 53 in (1.3 m) each
Shelf-Life: See use-by date on the Pad Cartridge [Section 7]
Accessories
Avive USB Power Adapter
Input Voltage 100-240 VAC 50-60 Hz
Output Voltage 5V DC 2A
Avive USB Charging Cable
Cable Length 3.3 ft (1 m)
Connectors USB Type A to USB Type C
Defibrillator Performance
Time to Shock Delivery following Patient Detection: 150J in less than 20 seconds
Patient Impedance Range: 25 Ohms to 200 Ohms
If the detected impedance is outside of this range, the user is prompted to check pad placement. The Avive AED
will not deliver a shock outside of this range.
Patient Analysis
The Avive AEDä measures the patient’s electrocardiogram (ECG) and uses a machine learning algorithm
[Rhythm Recognition Detector] to determine if defibrillation is appropriate. The Rhythm Recognition Detector
has been trained using numerous databases of adult and pediatric ECGs recorded from real patients. ECGs are
classified according to strict agreement criteria by independent cardiologists, and are divided into distinct sets
of data used for development and testing of the algorithm. A study entitled “Development and Validation of a