IFU-00367 Rev A
32
Immunity to RF Wireless Communications Equipment
Test
Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
Power
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380 390
TETRA 400
Pulse
modulation
b)
18 Hz
1.8
0.3
27
450
430 470
GMRS 460,
FRS 460
FM
c)
± 5 kHz deviation
1 kHz sine
2
0.3
28
710
704 787
LTE Band 13, 17
Pulse
modulation
b)
217 Hz
0.2
0.3
9
745
780
810
800 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
870
930
1720
1700 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
1845
1970
2450
2400 2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0.3
28
5240
5100 5800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0.2
0.3
9
5500
5785
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation,
it would be worst case.
Battery Safety UN38.3, IEC 62133, UL2054
Ingress Protection IEC60529, IP54
Shipping and Transportation ASTMD4169-16 Assurance Level I
Expected Service Life 7 years