5 Test Procedure
38 DHV0053 b2 EN
© b2electronic gmbh
5.4 Interrupting a Test
Once a test has started, it can be interrupted at any time. It is
recommended to select the appropriate method corresponding to the
situation
Situation Procedure
Routine STOP
(No emergency)
When a test is in progress, “STOP” on the
display screen is highlighted.
To interrupt the test, push in / click the
navigation knob (5)
• HVA28TD software deactivates HV
• Test stops
Alternative
When a test is in progress, press the HV
switch (6) to deactivate high voltage.
• HVA28TD hardware deactivates HV
• Test stops
Emergency Stop
In an emergency situation, press the
Emergency Off (1) to shutdown the system.
• HVA28TD hardware deactivates HV
• Test stops