PCA
5.7 Calibrate TA-Span
Material Required:
Thermocouple
Simulator
(K-type)
Range: 0 to 600 °F
Accuracy:
±0.5 °F
Procedure:
Calibration
1.
With the analyzer turned off, first plug the simulator’s
K-type connec-
tor
into the T-AIR jack (Figure 3-2); then turn on the analyzer and
wait for its warm-up cycle to complete.
2.
Enter the Calibration Menu Screen per Section 5.3; then choose
TA-SPAN to span the analyzer’s room-air/primary-air temperature
channel.
3.
Set the simulator to 212 °F (100 °C).*
4.
Wait until the MEASURED reading on the screen stabilizes. Then use
the
st
and
ENTER
keys to enter an APPLIED value that equals 212
°F
(100 °C).
Typical Calibrate TA-Span Screen During Calibration Procedure:
CALIBRATE TA-SPAN
MEASURED 209.0°F
APPLIED 0212.0°F
4.
Press the
RUN
key to calibrate the analyzer’s MEASURED value to
that
of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.
*
The TA-Span calibration range is 194 – 230 °F (90 – 110 °C). Any
attempt to calibrate outside this range will cause the analyzer to
display the message BAD CALIBRATION ENTRY.
Instruction 0024-9219
5-7