5. Advanced operation
2. Navigate to the Conguration > Calibration menu.
3. Enter the Display Function submenu.
4. Select one of the available display functions and conrm.
5.13 Amb ient Light Compensation (ALC)
About ALC
Ambient Light Compensation (ALC) can only be enabled on your display when the DI-
COM display function is selected. Therefore, please refer to "Display functions", page
23 to correctly set the display function.
When ALC is enabled, the DICOM display function will be recalculated taking a preset ambient light cor-
rection value into account. This value is determined by the selected reading room. Therefore, it is also
important to select a realistic reading room when enabling ALC. This can be done by following the instruc-
tions in "Reading rooms", page 24.
To enable/disable A LC:
1. Bring up the OSD main menu.
2. Navigate to the Conguration > Calibration > Ambient Light menu.
3. Enter the Ambient Light Compensation submenu.
4. Select Enabled/Disabled as desired and conrm.
5.14 Reading rooms
About reading rooms
Reading rooms can only be selected when the DICOM display function is selected.
Therefore, please refer to "Display functions", page 23 to correctly set the display
function.
The American Association of Physicists in Medicine (AAPM) composed a list of pre-dened reading rooms.
Each of these reading rooms are dened by following parameters:
the maximum light allowed in this type of room
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to
take into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is
enabled. Please refer to "Ambient Light Compensation (ALC)", page 24 to enable ALC.
The available reading rooms for your NIO 3MP LED Display are:
CR/DR/ MAMMO Corresponds to light conditions in diagnostic
reading rooms for computed radiology, digital
radiology or mammography. This setting has the
lowest maximum ambient light.
CT/MR/NM Corresponds to light conditions in diagnostic
reading rooms for computed tomography, magnetic
resonance or nuclear medicine scans.
24 K5902067 NIO 3MP LED DISPLAY 10/04/2013

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Barco MDNC-3321 and is the answer not in the manual?

Barco MDNC-3321 Specifications

General IconGeneral
BrandBarco
ModelMDNC-3321
CategoryMonitor
LanguageEnglish

Related product manuals