41394 Rev. A
July 2017
Preventive
Maintenance
SIGMA SPECTRUM INFUSION PUMP WITH
MASTER DRUG LIBRARY
35700BAX
Pump Operating Software Version 6.02 and 6.05
For use with Master Drug Library Version 6
Question and Answer IconNeed help?

Do you have a question about the baxter 35700BAX and is the answer not in the manual?

baxter 35700BAX Specifications

General IconGeneral
Part Number35700BAX
Power SourceElectric
MaterialCast Iron
Model35700BAX
Voltage115V

Summary

Preventive Maintenance

Preventive Maintenance and Frequency

Defines PM, requirements, and recommended frequency for pump maintenance.

Pump Identification

Serial Number and Operating Software Version

How to find and record the pump's serial number and software version.

Network Connectivity

Steps to verify the pump's current network connectivity status.

Active Drug Library

Visual Inspection

Visual Inspection - AC Power Adaptor

Equipment Required

Keypad Operation Test

Memory Test

Pump Operation Tests

Test Setup

Prepares the pump for operational tests by setting up IV sets and fluid.

Downstream (Distal) Occlusion Sensor Test

Tests the downstream occlusion sensor's functionality and pressure limits.

Upstream (Proximal) Occlusion Sensor Test

Tests the upstream occlusion sensor's functionality and alarm activation.

Flow Rate Accuracy Test

Measures flow rate accuracy using volumetric or gravimetric methods.

Air Detection Test

Verifies the pump's ability to detect and alarm for air in the line.