Preventive Maintenance 3
SIGMA Spectrum Infusion System Pump Operating Software Version 6.02 and 6.05
Preventive Maintenance For Use With MDL Version 6
2. Select Biomed Options and press OK.
3. Enter the Biomed password and press OK.
4. Select Network Configuration and press OK.
5. Select Network Status and press OK.
When using a WBM, a wireless icon (in the top right corner of the Pump display) will indicate
the status of the network connection. See the “Pump Icons” section of the device’s Operator’s
manual.
6. Record the status on the Preventive Maintenance Check Sheet.
Active Drug Library
If using a WBM, the Drug Library will load automatically from the Gateway Server once the
Network Configuration is complete.
NOTE: It may take several minutes for the Drug Library BDL file to begin downloading to the
Pump.
The message, “Activating new drug library” appears with a progress bar on the Pump’s display.
To confirm Drug Library activation:
» When the new Drug Library has been activated on the Pump, the screen will display
“COMPLETE”. Press OK to confirm activation.
NOTE: If the Drug Library is not current or is missing, see “Programming the Pump” in the
device’s Operator’s Manual.
WARNING Confirm Drug Library.
Master Drug Library Administrators (MDLAs) should verify the
correct Drug Library is installed when deploying the Drug Library to
Pumps.
Master Drug Library Administrators (MDLAs) should verify the
Drug Library transfer is successful after deployment.
Users should verify the correct Drug Library is installed on the
deployed Pumps.
Before implementation, clinical users at each facility must
thoroughly test and validate their Drug Library per their facility’s
procedure to ensure configuration and workflow reflect clinical
practice.
To identify the Active Drug Library:
The name and version of the Pump’s active Drug Library BDL file can be viewed on two screens of
the Pump’s display: the Power Status screen and the User Options screen.
Record and verify the Pump’s active Drug Library BDL file using one of the following methods.