Preventive Maintenance 9
SIGMA Spectrum Infusion System Pump Operating Software Version 6.02 and 6.05
Preventive Maintenance For Use With MDL Version 6
Memory Test
To test the Pump’s memory:
1. Plug the AC Power Adaptor into a working wall outlet, or install a Battery Module.
2. Press ON/OFF to power on the Pump.
3. Select any Care Area and press OK.
NOTE: If the “New Patient Yes/No?” prompt appears, press the yes soft key.
4. Press OK to view the Drug Library.
5. From the Drug Library, select BASIC and press OK.
6. Select Primary Bag and press OK.
7. For Mode select mL/hr and press OK.
8. Set the Rate to 100 mL/hr and press OK.
9. Set VTBI mL to 50 mL and press OK.
10. Press ON/OFF to power off the Pump.
11. Press ON/OFF to power on the Pump.
12. On the “New Patient Yes/No?” screen, press the no soft key.
13. Verify that the Rate and VTBI parameters were saved. (100 mL / 50 mL)
14. Press ON/OFF to power off the Pump.
15. Record the findings on the Preventive Maintenance Check Sheet.