Preventive Maintenance 21
SIGMA Spectrum Infusion System Pump Operating Software Version 6.02 and 6.05
Preventive Maintenance For Use With MDL Version 6
Pump Inspection
Inspection Pass Fail
1. Visual Inspection - Pump
Inspect the Pump for visual signs of damage and damage to all exterior components such as
dents, cracks, cuts, heat damage or fluid intrusion.
Pump Casing
Pump Display
Keypad
Keyhole
Pump Door
Power Connector on back of rear case
IV Set Tube Channel
Verify legibility of all labels. Replace labels as necessary.
2. Visual Inspection - AC Power Adaptor
Inspect the AC Power Adaptor for visual signs of damage and damage such as cracks, bent
prongs, cuts, exposed wires or dents.
AC Power Adaptor Module
AC Power Adaptor Cord
AC Power Adaptor Cord Retainer
AC Power Adaptor Module LED
Verify legibility of all labels. Replace labels as necessary.
3. Network Configuration - Wireless Battery Module
Verify that the network configuration file is installed and that there is
connectivity (green Icon).
If no network configuration file is installed, see “Wireless Networking” in
Service Manual.