Preventive Maintenance 1
SIGMA Spectrum Infusion System Pump Operating Software Version 6.02 and 6.05
Preventive Maintenance For Use With MDL Version 6
PREVENTIVE MAINTENANCE
Read each PM test completely prior to performing.
To ensure that the SIGMA Spectrum Infusion System is operating within specified parameters,
perform the PM tests in this section according to the maintenance schedule below.
After each operation in this procedure, record the results on a copy of the “Preventive
Maintenance Check Sheet” on page 20.
NOTE: SIGMA Spectrum Technical Training is available, but not required, for performing Pump
PM. Contact MedinaTechTraining@baxter.com for more information.
Preventive Maintenance and Frequency
PM consists of:
Scheduled inspections, testing, and performance evaluations
Servicing any SIGMA Spectrum Infusion Pump that fails any inspection, test, or performance
evaluation
PM is required on:
All Pumps in service (use)
Any Pump suspected of being physically damaged, dropped, or having fluid intrusion.
Recommended PM frequency:
Pump: Annually
NOTE: A PM or network expiration due date may be entered in the BIOMED options. A “Due for
Inspection” reminder will appear at power on followed by the programming screens. At
power off, a “Due for Inspection” display will appear prior to Pump shutdown.
When performing an annual PM on SIGMA Spectrum Infusion Systems, record the findings on the
check sheet provided. See “Preventive Maintenance Check Sheet” on page 20.