4 Preventive Maintenance
Manual 41394 SIGMA Spectrum Infusion System
Revision A Preventive Maintenance
The Power Status screen (the Pump must be powered off):
1. Press ON/OFF to power on the Pump and view the current software version and serial
number.
2. Record the Pump’s active Drug Library BDL file on the Preventive Maintenance Check Sheet.
3. Verify the current Drug Library BDL file in use with the facility’s pharmacy or MDLA.
4. Record the facility’s current Drug Library BDL file on the Preventive Maintenance Check Sheet
and verify they are the same.
5. Press ON/OFF to power off the Pump.
NOTE: The User Options screen (the Pump must be powered off):
1. Press ON/OFF to power on the Pump and view the current software version and serial
number.
2. From the Select Care Area screen, press the options menu soft key (software version 6.05)
or Help then the Options soft Key (software version 6.02).
3. Select User Options and press OK.
4. Select View Information and press OK.
5. Select Library Information and press OK.
6. Record the Pump’s active Drug Library BDL file and date on the Preventive Maintenance
Check Sheet.
7. Verify the current Drug Library BDL file in use with the facility’s pharmacy or MDLA.
8. Record the facility’s current Drug Library BDL file on the Preventive Maintenance Check
Sheet.
9. Press ON/OFF to power off the Pump.
Visual Inspection
To inspect the Pump:
1. Inspect the Pump for visual evidence of damage and/or defects to exterior
components such as
dents, cracks, cuts, heat damage or fluid intrusion.