ENUSER’S MANUAL RTI
48
27. AUTO-OFF
To program the function of auto switch off, press the key during the initial count. On the screen appears the message AUTO-
off 0,1,5,30
0= disabled
1= the balance is going to switch off after 1 minute of not using it
5= the balance is going to switch off after 5 minutes of not using it
30= the balance is going to switch off after 30 minutes of not using it
Choose the options with the key
and then validate with the key .
The balance is going to go back to its usual weighing mode.
28. MEMORY CLEAR
This procedure clears the registers of accumulations from the memory of the equipment. It is recommended to do a memory clear
after printing correctly all the reports needed.
Press the key for 3 seconds, on the screen appears the message “clear all record” “Are you sure”, press the key to
confirm or to go back to the weighing mode without clearing anything.
PROGRAMMING MANUAL
29. PARAMETERS
To access the parameters, press the key when doing the initial count. The balance is going to ask for the PIN ( Pn) . Write
0000 with the numerical keyboard and confirm the operation with the key . Use the key to scroll between the two
menus, use the key to return to the weighing mode.
!
NOTE: to access the parameters marked with an asterisk it is necessary to press the switch placed
under the keyboard, in the inferior part of the equipment before pressing the key to access. This
one is protected by a seal. Breaking the seal the verification is lost.
DISPLAY DESCRIPTION
(*)
CAL F0
GAUGING
When the parameters menu shows F0 the user must press the key
to access.
On the screen will appear the message UNLOAD, which indicates that the weight must be removed from
the platform.
When the indicator is steady, press the key
to program the cero point.
The screen is going to show then the required calibration weight. Press the key
to accept.
LOn the screen is going to appear the message LOAD. Place the weight to calibrate on the balance. When
the indicator is steady, press the key
. Now the balance starts the calibration process.