Operating the bbe BenthoTorch
bbe BenthoTorch Manual Page 20
green:
0.7µg/cm’2
more <- -> ESC
cyanos:
0.7µg/cm’2
more <- -> ESC
diatoms:
0.7µg/cm’2
more <- -> ESC
More switches between the following items:
Automatic alternating of the tree algae classes
Display of the concentration of cyanobacteria
Display of the concentration of green algae
Display of the concentration of diatom
Display of the time / date of the measurement.
Use the arrow keys to display data of another point in time.
Press “ESC” to return to the start measurement item.
I
NTERVAL
/
CONTINUOUS MEASUREMENT
Start the BenthoTorch
Use a carrying cable/rope to secure the instrument
Ensure that the connector is closed with the cap / dummy connector
Do not exceed the maximum depth (10 m / 100 m).
Use the continuous mode (settings -> mode -> continuous)
Adjust measuring time and interval
Go to “Start -> Start continuous measurement”
Apply the BenthoTorch in to the benthic algae. Ensure that the red foam shadows the algae from the light.
To stop the BenthoTorch, turn it upside down and back between two measurements when the BenthoTorch
switches off because of the selected measuring interval. If the measuring interval is 0 s, use the exit key
between two measurements:
To read the data, go to “Data -> view
Reference method for chlorophyll-a measurement for validation of the bbe BenthoTorch
S
UBSTRATE USED FOR REFERENCE
For test purposes, substrata used are exposed, flat materials which are non-fluorescent in visible light and easy
to scrape:
a. Black PVC plates (4x6cm)
b. Gray PVC plates (4,5x7,5cm)
c. Glass microscopic slides darkened on one side to omit the “double sidedness problem”.