evoStim
®
E - User guide 19
the stimulation level it is required to start the session
8.1.6) otherwise stimulation stops after a few
seconds.
8.1.6 Start the session
Briefly push the button to start the session.
8.1.7 Adjust the stimulation level during the
session
10 seconds after starting the session, all the "touch"
commands will be disabled (except the "Buzzer
Enable / Disable" and the backlight adjusting) to
prevent inadvertent changes; the symbol (⑮) will
appear on the display.
Turn for at least 1/2 a turn the knob , clockwise,
to temporarily unlock the commands, touch the area
, then turn the knob ①6clockwise (to increase) or
counter-clockwise (to decrease) until you feel a
strong and comfortable level of stimulation. If
stimulation is too strong, turn the knob
counterclockwise.
8.1.8 - Conclusion of the session
At the end of the session, make sure that the device
is turned off and then gently pull the probe out,
avoiding to pull on the wires, as this could damage
them over time. Rinse the probe thoroughly under
tap water, carefully dry and store in the original bag.
8.2 - OPTIONAL OPERATIONS
8.2.1 Temporarily stopping the session
(PAUSE)
During the session, you can temporarily pause the
programme to modify a parameter.
Briefly press the button to pause the session.
The pause state is visually reported by the symbol
and the green back-light of the display
flashing (if enabled). In the PAUSE state, you can
change: the session time, the frequency, the pulse
width, the Action time, the Rest time (see § 8.2 for
details). To resume the session, briefly press the
button ①.