evoStim
®
E - User guide 31
11
Technical features
Electrostimulation output: 1 to 99 mA
pp
in steps of
1 on a standard load 1KΩ. (with pulse-width of 200
µs).
Frequency: adjustable from 1 to 150 Hz 5% of
Full-Scale).
Frequency during rest: - 3 to 5 Hz. (± 5% of FS).
Pulse Width: automatically adjusted between 50
and 400 µs in steps of 10µs (± 5% of FS).
Output wave-shape: selectable between “bi-phasic
symmetrical” pulses and “bi-phasic symmetrical
alternated”.
Rise/Fall time: 0,3 to 1s. (automatically adjusted).
PLATEAU time - 1 to 60 s. In steps of 1 s. (± 0,1 s.)
REST time - 0 to 60 s. In steps of 1 s. (± 0,1 s.).
Session time: adjustable between 5 and 90 min’ in
steps of 5 min’ (± 1 s.) or Continuous.
Biofeedback scale (FS): selectable among
2-4-8-12-16-20-24-32-40-50-60-70-80-90-100
-120-140-160-180-199 µV.
Supply voltage 4,5 V by 3 alkaline batteries 1,5
Volt type AAA (LR03).
Battery life: 20 hours average (variable according
to the program and the output level set).
Electrical safety: Internal supply according to CEI
62-5 IEC 601-1 (3a edition).
Overall dimensions: 73 x 147 x 25 mm.
Weight: About Kg 0,2 (including batteries)
Environment using limits: +5 / +40°C U.R.15% /
93% 700hPa / 1060hPa.
Transport and storage limits:: -25 / +70°C U.R.
93%.
Electromagnetic susceptibility: Class A according
to CISPR 11.
Patient circuit connection: 1 channel (Ch.1) with
micro axial connector to which connect the supplied
gray bipolar cable ending with 2mm protected plugs.
- 1 input (EMG-Ref.) with axial micro connector to
which to connect the single-pole blue cable, ending
with a 2mm protected plug.