BEACMED evoStim E - Trouble-shooting Table

BEACMED evoStim E
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
evoStim
®
E - User guide 33
12
Trouble-!
shooting table
Sign
Comments /
Probable causes
Suggested remedies
12.1 - The unit does
not switch-ON
12.1a - Check the batteries are inserted correctly (§ 7.1).
12.1b - Batteries are low.
12.1bb - Replace the batteries with new ones (§ 7.1).
12.1c - Make sure that the
contacts of the battery are not
missing, broken or oxidised.
12.1cc - If batteries had remained for a long time into an
unused unit, some battery acid may leak onto the contact
springs. Try to clean the contacts and replace batteries with
new ones. Otherwise contact the manufacturer.
12.2 - The unit will
not switch-OFF
12.2a - Be sure to properly press
the shutdown button and hold it
for at least 2 seconds (§ 8.1.2).
12.2aa - Remove the batteries from the battery compartment
and wait at least one minute before reinserting them (§ 7.1).
12.3 - Dopo
l'accensione
l'apparecchio si
spegne da solo.
12.3a - If the unit is switched on, but a programme not started within 5 minutes, the unit will
automatically switch off to conserve battery power.
12.3b - The batteries are low.
12.3bb - Replace the batteries with new ones (§ 7.1).