34
BEACMED s.r.l. - Via Monte Bianco, 12 - 27040 Portalbera PV (Italia)
Sign
Comments /
Probable causes
Suggested remedies
12.4 - You can not
adjust the intensity
level of electro-
stimulation.
12.4a - Be sure to carefully follow the steps specified in the user manual (§ 8.1.5 - § 8.1.6).
12.4b - The device is equipped
with a Control Lock system to
prevent accidentally changing the
intensity level and the parameters
during the session (§ 8.1.7).
12.4bb - To unlock the controls and allow the intensity
adjustment during a treatment session, make a clockwise
rotation of the knob fully clockwise and then touch the intensity
value, it is now possible to change the value. The control-lock
system will return after 5 seconds of inactivity on the
operator's controls of the unit.
12.4c - When you try to set the
intensity, the value drops to "0"
and the display will flash red.
12.4cc - The output circuit connection to the patient is
interrupted. Check the connection of cables and electrodes. If
required, replace the cable/s. If the wires and their connection
is ok, there may be insufficient contact between the electrodes
and the skin. Try to moisten the gel surface of the electrodes
with water. If necessary, replace the electrodes with new ones.
12.5 - The unit is
ON but no
operations are
allowed.
12.5a - Remove the batteries,
wait a few minutes and re-insert
them.
12.5aa - If the problem persists, contact the manufacturer or
distributor.