8.2.2 Changing the session time (min’)*
Tap the area ②, the two digits flash, then rotate the
knob ①66to the desired time.
8.2.3 - Modify the frequency (Hz)
Tap 2 times the area ②, then rotate the knob to the
desired value. The frequency can only be adjusted
before starting the session (§ 8.1.6).
8.2.4 - Display the pulse-width (µs)*
To view the pulse-width (µs), tap 3 times the ②
area, the pulse-width can not be changed.
8.2.5 - Modify the ACTION! time (sec)*
Tap the area ③,%%it will blink for a few seconds, ②
the display ②% will shows the current value of the
action time (sec.). To edit it, touch area and ②,
while the display is flashing, turn the knob to obtain
the desired value.
8.2.6 - Modify the RELAX time (sec) *
Tap the area ④,% %it will blink for a few seconds, ②
the display ②% will shows the current value of the
action time (sec.), To edit it, touch area and ②,
while the display is flashing, turn the knob to obtain
the desired value.
8.2.7 Manual adjustment of the biofeedback
scale *
If the EVO or BIOFEEDBACK program has been
selected, requiring the detection of the EMG signal,
it is possible to adjust its sensitivity, that is the Full-
Scale (FS).
Tap the area , the 3 digits will blink for few
seconds, during which it is possible to change the
scale by turning the knob.
8.2.8 Automatic adjustment of the
biofeedback scale *
Alternatively, the Full-Scale (FS) or detection
sensitivity can be automatically adjusted. With the
probe connected (§ 7.3) and suitably inserted (§
7.4), tap the area , the 3 digits will blink for few
seconds, tap the centre of the Concentric Circles
Target (CCT), the 3 digits will show now the
pressure value acquired by the balloon probe.